JG515 type simulated wearable pre-processing machine
【Scope of application】
The simulated wear pretreatment machine is used to detect the simulated wear pretreatment of the respirator before other tests such as breathing resistance, filter material penetration, etc.
【Standards compliant】
EN149-2001 8.3 etc.
【working principle】
The simulated wearable preconditioner is composed of a breathing simulator, a test head model, a saturated water vapor generator, a temperature control device, a flow meter, etc. It can control the breathing simulator to run at 25 breaths per minute, with a breathing volume of 2L each time. The temperature of saturated water vapor near the mouth of the head die can be controlled to (37+2) °. Turn on the pretreatment device to make the breathing device and the saturated water vapor generator enter the working state until it reaches a stable state. Put the exhaler on the matching test head model in an airtight manner, and test for about 20 minutes. Remove the particulate filter material from the respirator, then reinstall it, and continue to repeat the test 10 times.